# Setup for live cell
|
|
|
## 1. Turn on the main CO$_2$ supply
|
|
|
The carbon dioxide is supplied from the tank opposite the Nikon Ti room. When CO$_2$ is not being used the tanks should all be off and the main gauge reading 0 kPa.
|
|
|
|
|
|
Open the active tank by turning the tap on the top of the tank. The main gauge should now read about 400 kPa.
|
|
|
![](media/CO2_tanks_50.jpg)
|
|
|
|
|
|
|
|
|
|
|
|
## 2. Turn the CO$_2$ regulator in the room
|
|
|
The CO$_2$ is supplied to the incubator from the tap in the wall behind the Nikon Ti.
|
|
|
|
|
|
* Open blue tap fully
|
|
|
* Open the regulator. The gauge should read 20-40.
|
|
|
|
|
|
![](media/CO2_wall_10.png)
|
|
|
|
|
|
## 3. Check the water level in the humidifier bottle
|
|
|
There is a bottle inside the microscope cage on the right hand side below the stage that humidifies the air coming from incubator. Check that this half full and top up if necessary.
|
|
|
|
|
|
|
|
|
## 3. Turn on the Incubator
|
|
|
|
|
|
* Turn on the incubator on from the blue switch
|
|
|
* Make sure the flow rate dial is on. You should see bubbling in the humidifier bottle.
|
|
|
* Set the target temperature and CO$_2$ concentration by pressing the green/red buttons to increase/decrease the targets.
|
|
|
|
|
|
![](media/incubator_markup_20.jpeg)
|
|
|
|
|
|
|
|
|
## 5. Attach the well plate cover
|
|
|
The well plate cover should be in the microscope cage attached to the humidifier bottle.
|
|
|
|
|
|
This sits on top our well plate on the stage
|
|
|
![](media/wellplate_cover_markup_20.jpeg)
|
|
|
![](media/wellplate_on_stage_17.6.jpeg)
|
|
|
|
|
|
|
|
|
## 5. Wait for equilibrium
|
|
|
|
|
|
It will take some time for the conditions in sample to reach their target values.
|
|
|
|
|
|
* **CO$_2$ - 10 min**: The air above the sample will equilibrate very quickly, less than 5min.
|
|
|
* **Temperature - 1 hour**: The temperature of the sample will take longer as large volume of air in the cage needs come to equilibrium. For a 15°C difference from room temperature this takes about an hour. If you need higher temperature you will need to wait longer - check current temperature in the cage on the incubator screen.
|
|
|
* **Stage movements - 2 hours**: The microscope stage expands it warms up which shifts the sample slightly. Changes in Z can be corrected for using the PFS system, but the shift in XY position may affect your experiment. For high-resolution time-lapse imaging allow 2 hours for equilibration.
|
|
|
|
|
|
![](temp_disp.png)
|
|
|
|
|
|
# Shut down for live cell
|
|
|
1. Turn off the incubator.
|
|
|
2. Place well plate cover next to the humidifier bottle.
|
|
|
3. Top up the humidifier bottle.
|
|
|
3. Turn the regulator and tap off at the wall.
|
|
|
4. Turn the tap on the CO$_2$ tank off |
|
|