Leak Air table on the Deltavision-OMX-SR
2018-02-09
Problem: Air table not floating
Symptoms: Hissing sound
Action: Contact Louise Cole or Mike Johnson who will resolve the issue. If unresolved by solution below MIF officers will log an incident and request repair.
Solution: Tubing to air table was disconnected. To reconnect, see photos below:
Steps to fix
1. Open back cover (top RHS, at back)
2. Open front cover (top RHS, at front).
Carefully move glass bottles out of the way. (Disconnect tubing to glass bottles if needed but be sure to replace afterwards).
3. Locate Air Table tubing
this is clearly labelled and reconnect.
4. Check air table is now floating
Replace glass bottles and covers as shown.