Skip to content
GitLab
  • Menu
Projects Groups Snippets
  • /
  • Help
    • Help
    • Support
    • Community forum
    • Submit feedback
    • Contribute to GitLab
  • Sign in
  • M Microscope-instructions
  • Project information
    • Project information
    • Activity
    • Labels
    • Members
  • Repository
    • Repository
    • Files
    • Commits
    • Branches
    • Tags
    • Contributors
    • Graph
    • Compare
  • Issues 0
    • Issues 0
    • List
    • Boards
    • Service Desk
    • Milestones
  • Merge requests 0
    • Merge requests 0
  • CI/CD
    • CI/CD
    • Pipelines
    • Jobs
    • Schedules
  • Deployments
    • Deployments
    • Environments
    • Releases
  • Monitor
    • Monitor
    • Incidents
  • Analytics
    • Analytics
    • Value stream
    • CI/CD
    • Repository
  • Wiki
    • Wiki
  • Snippets
    • Snippets
  • Activity
  • Graph
  • Create a new issue
  • Jobs
  • Commits
  • Issue Boards
Collapse sidebar
  • MIF
  • Microscope-instructions
  • Wiki
  • Dv_omx_sr
  • Trouble shooting
  • Air table leak
  • air table leak

air table leak · Changes

Page history
Added air table leak, extracted from docx authored Feb 12, 2018 by 119898's avatar 119898
Hide whitespace changes
Inline Side-by-side
DV_OMX_SR/trouble-shooting/air-table-leak/air-table-leak.md 0 → 100644
View page @ 9b90a02a
# Leak Air table on the Deltavision-OMX-SR
Feb–9^th^ 2018
**Problem:** Air table not floating
**Symptoms:** Hissing sound
**Action:** Contact Louise Cole or Mike Johnson who will resolve the
issue. If unresolved by solution below MIF officers will log an incident
and request repair.
**Solution:** Tubing to air table was disconnected. To reconnect, see
photos below:
## Steps to fix
### 1. Open back cover (top RHS, at back)
![](./media/image1_40.jpeg)
### 2. Open front cover (top RHS, at front).
Carefully move glass bottles
out of the way. (Disconnect tubing to glass bottles if needed but be
sure to replace afterwards).
![](./media/image2_40.jpeg)
![](./media/image3_40.jpeg)
### 3. Locate Air Table tubing
this is clearly labelled and reconnect.
![](./media/image4_40.jpeg)
![](./media/image5_40.jpeg)
### 4. Check air table is now floating
Replace glass bottles and covers as shown.
![](./media/image6_40.jpeg)
![](./media//media/image7_40.jpeg)
Clone repository
  • Booking System
    • ACLS
      • ACLS
  • DV_OMX_SR
    • DV OMX SR spec
    • DeltaVision OMX SR
    • trouble shooting
      • air table leak
        • air table leak
      • trouble shooting
  • Fiji
    • Fiji install
    • Fiji
    • Shortcuts
    • timelapse to movie
      • Making a movie
  • Filters
    • BGRFR
      • BGRFR
      • Di
      • Em
      • Ex
    • Filter List
    • TexasRed
      • 3100_Di_600 660
View All Pages