# Setting up Koehler Illumination
|
|
|
|
|
|
![](condenser-markup.jpg)
|
|
|
|
|
|
1. Select the lens
|
|
|
2. Click on the Bright Field OC (or set the condenser to `A`)
|
|
|
3. Make sure the polariser is out
|
|
|
4. Focus on the sample, the edge of the slide or coverslip is easy place to start
|
|
|
5. Close the Condenser Aperture Diaphragm, you should see the octagon formed by the blades
|
|
|
6. Raise and lower the condenser so that the blades are in focus
|
|
|
7. Adjust the centering screws so that the octagon in the centre of the image
|
|
|
8. Open the Condenser Aperture Diaphragm so the blades are just past the camera field of view
|
|
|
9. Remove eye piece and adjust the Condenser Aperture so that 2/3 of pupil is filled.
|
|
|
|
|
|
## For Phase Contrast
|
|
|
|
|
|
1. Click on the lens OC (or set the condenser to correct Phase setting)
|
|
|
2. Remove the eye piece and
|
|
|
|
|
|
## For DIC
|
|
|
|
|
|
1. Insert the Polariser (or set the condenser to 'A')
|
|
|
2. Click on the lens OC
|
|
|
3. Adjust the angle with silver rod to give the best contrast. |
|
|
\ No newline at end of file |